LE Series EO sterilizer
The brand, quantity of parts , function, control system , overall standard of BE, LE and GE series are different, so are the costs.
These three series are different. Customers can choose one series according to functional requirements.
-GE series has the highest price because of its high configuration, full functions and the most detailed reports.
-LE series, can meet the needs of most customers, low price.
-BE series, the lowest price, suitable for customers with less capital budget.
Description Relevant Technial Parameter of LE Series
system design standard: EN1422. ISO11135-2014
body (main part) LE Series EO sterilizer
1. Cabinet:stainless steel SUS304,δ=6mm
2. water jacket: U-bar carbon steel (bottom is stainless steel)
3. housing:stainless steel
4. foot: U- bar steel 14#/Q235
5. foot support: steel sheet 10/Q235
6. insulating layer: ceramic fiber
door (main part) LE Series EO sterilizer
1. door: stainless steel : SUS304,δ=6mm
2. door frame:rectangular steel 120*60/Q235
3. door water Jacket: U- bar carbon steel
4. door seal flange: carbon Steel
5. housing:stainless steel
6. insulating layer:ceramic fiber
heating components (main part) LE Series EO sterilizer
1. frame:angle steel 50×50×3
2. heating water tank: stainless steel
3. housing of tank: stainless Steel
4. electronic heater: stainless steel electrical heating pipe
EO vaporizer
1. vaporizer liner:stainless steel
2. EO coil pipe:stainless steel pipe SUS304
3. vaporizer housing:stainless steel
4. insulating layer:fireproof aluminium silicate fiber blanket
EO feeding device pressure gage,mannual valve , Filter , connecting pipe
Standard Components LE Series EO sterilizer
Description Specification & ParameterBrand
-
heating pipe: 3*3kw/AC380V 3P
-
vaccum pump: 1.45kw/AC380V 3P 50HZ
-
vaporization pump: AC220V/200W 2P 50HZ
-
safety valve: 0.15Mpa1
-
steam generator: 3KW/AC380V 50HZ
-
air compressor: 1.5Kw/AC380V 3P 50HZ
-
magnetic valve DC24V
-
pneumatic valve EO sterilizer
-
humidifying valve: 20mm screwed fitting
-
vacuum valve: 1inch screwed fitting
-
Feeding EO valve:20mm screwed fitting
-
Feeding N2 valve:none
-
air release valve:1 inch screwed fitting
-
steam heating valve: none
EO electric scale
1. range: 100Kg;
2. material: stainless steel;
3. support RS485 communication function;
Control System LE Series Description Specifications
temperature sensor
output form:PT100;
range:-50~100℃
pressure sensor
output form:4~20mA
voltage: DC24V
range:-0.1~+0.15Mpa
humidity sensor
output form:4~20mA
voltage: DC24V
range: 0~100%RH
control cabinet: integrated work station BGIB2-E01
control module: SIEMENS S7200 PLC/SMART
screen: TK6050IP
low voltage apparatus: SCHNEIDER
buttons and indicator light: SCHNEIDER
IPC(Industrial Personnal Computer): SMART IPC3000
screen: 226V6QSB6/21'5/16:9
printer: Deskjet 2020hc Color Ink-Jek
UPS: 1000VA SANTE
System Function LE Series
-
system design standard: EN1422. ISO11135-2014
-
system control mode: by switching. Soft mannual & Full- Automatic
-
open/close Door way: pneumatic revolve door
-
door qty: one
-
data record way: interval mode , warning triggered mode
-
historical curve record way: continuous time interval mode
-
data save time: 10 years(If hard drive is not damaged)
-
recirculated water connector: boht cabinet and door with one way in and one way out
-
circulation water jacket replacement time: <5 minutes
-
temperature sensor precision: display Range: 0-100℃;resolution: 0.1℃
-
temperature control precision: cabinet wall temperature difference within ±5℃, cabinet inside temperature difference is within ±3℃
-
pressure sensor precisiondisplay: Range : -100-+150Kpa,resolution: 0.1Kpa
-
vaccum control mode: continuous mode and stepping mode ( proportional valve regulating mode is choosable)
-
vaccuming time: <30minutes from 0 to -90KPa
-
humidity sensor precision: display range: 0-100%RH(no-condensing) , resolution 0.1%RH
-
control time range: 0-99hours, sterilizing time resolution 1min
-
cabinet water volume auto detecting yes ,system shows and controls the water volume and feeds water automaticly.
-
steam heating connector: yes, auto control
-
steam detecting: yes, displaying and warning when the steam is not enough
-
vaccum pump detecting connector: yes, displaying and warning when the pressure is not enough
-
steam pressure detecting connector: yes, displaying and warning when the pressure is not enough
-
humidifing connector: yes ,humidifying automaticly by pulsatile mode
-
humidifying way: pressure difference or sensor
-
Nitrogen protection system connector: yes , auto control
-
tail gas treatment connector: yes , auto control
-
in-door and out-door air exhausing connector: yes , auto control
-
data record and dealing mode: PC monitoring software and sterilizing Data report software
-
over high temperature limit protection: yes
-
over high pressure limit protection: yes
-
EO under low temperature limit protection: Yes
BOCON is a large-scale manufacturer of ethylene oxide sterilizer. We can customize and produce ethylene oxide sterilizers to meet all the needs of customers.
Jul 08, 2025
view: 1858
This article introduces the temperature, ventilation, explosion-proof and safety requirements for sterilization workshops....
Read More
Jun 17, 2025
view: 1928
The sterilizer is used for biological control as a channel between areas. Usually, in the environment with the highest sealing requirements, rooms with double doors are sealed on the wall, forming a barrier between different categories of areas. Cros...
Read More